1. Connect one end of the 0.7MΩ standard resistor to the ground wire of the withstand voltage instrument.
2. Turn on the power, and set the instrument and alarm leakage current at 5mA.
3. Turn on the instrument, hit the other end of the standard resistance with a test rod, and adjust the voltage. If the instrument sends out an alarm within the range of 3410V to 3590V, it is determined that the instrument is in a normal working state. If the instrument alarms within the range of 3410V to 3590V, the instrument is not working properly .
4. When the function of the equipment is found to be invalid during the operation inspection, and the operation inspection results cannot meet the specified requirements, the operator needs to re-test the products that have been tested since the last operation inspection passed, and send the instrument for maintenance.